This page has been moved. If you are not redirected within 3 seconds, click here to go to the Mesalabs home page.
HDU-Pt100-I
High Accuracy Temperature Sensor
Interface
HDU-Pt100-I
High accuracy
Different sensors with
high accuracy available
Interfaces which are
supplied including sensors are
adjusted to each other and have the high system
accuracy specified in the technical data below
Technical Data
Temperature HDU-Pt100
Range:
0...100 °C
Resolution:
0.001 °C
Accuracy:
25 ... 40°C ± 0.1
°C
otherwise ± 0.2 °C
if Sensor and interface
are adjusted together
Calibration
PTB/NIST traceable
Communication rate
Five
times per second
Degree of
protection
IP54
Power supply
3,3 to 5 V, max. 70 mA
via RS485- or USB-Interface
Environmental
conditions
Ambient Operating Temperature Range: 15 to
35°C
Storage Temperature Range: -10 to 60°C
Operating and Storage Humidity: 85% max
relative humidity, non-condensing