What is the calibration period of IBP sensors and meters?

The calibration period is one year. We recommend to double check the conductivity once per month with a
conductivity reference solution. If biological material grows inside the cell, it could interfere the accuracy
of the conductivity measurement.

We know that competing companies do not recommend a monthly test of the conductivity measurement,
some even offer this as a "feature" of their meters. From our more than 25 years experience in developing
dialysis meters we strongly recommend monthly testing as a safeguard precaution.
 


     
  If I send my sensor or meter in for calibration, what must I included in my shipment?

As the conductivity electrode is part of the calibration process, it must be included with the meter.
If your meter has a flow probe, it must also be included with your shipment. There is no need to
ship the battery charger or other accessories.
 


     
  Why do IBP Conductivity Sensors allow you to select the Temperature Coefficient of the Dialysis Machine to be calibrated?

This is a frequently misunderstood topic that is often neglected in hemodialysis.
Click here to read the background.
PDF (23 kByte)
 


 
   
 
     
  Home Products Company Support eMail us Address  
  SmartHDM About us Sales & Service Mail us IBP Medical GmbH
Ikarusallee 15
30179 Hannover
Germany
     
+49 511 957 336 0
 
 
    Dialysis Meters Privacy Statement FAQ We want your feedback
    HDU-Sensors QM-Certificate Meter Pick Up Service Sign our Newsletter
    Calibration Service Imprint Terms & Conditions  
        Warranty  
        Worldwide Distributors  
        Online Invoice Payment  
 
 
Copyright 2022 IBP Medical GmbH IBP is a subsidiary of Mesa Laboratories, Inc. back to top