IBP Instruments Service is a cornerstone in our philosophy Hemodialysis, Meters for Calibration - Chairs
  Home   |   Products   |        |   Contacts  |   Help Center  |  eMail us |  Home
  Reference Meters  Calibration Service  |  Heat Exchanger  
 
Reference Meters

 
Comparison IBP and Mesa meters
    MESALabs MESALabs IBP IBP  
        HDM97BP HDM99XP  
      NEO2 90XL HDM97Pocket HDM99XP  
Measuring Number of measuring channels   4 4 4 9  
  Conductivity    
  Temperature    
  Pressure    
  pH    
  Flow        
  Voltage          
  Frequency/Pulses/Events          
Features Graphic on screen          
  Oscilloscope Function          
  Temperature Compensation
selectable by brand name
       
  Automated Pressure Drop Measurement          
  Computer Interface        
  Data Acquisition Software        
  Warranty   12 month 12 month 24 month 24 month
               
Measuring Conductivity accuracy (2 ... 20 mS/cm)   ±0.1 mS/cm * ±0.1 % ** ±0.03 mS/cm ±0.03 mS/cm
  Temperature accuracy (25 ... 40°C)   ±0.2 °C * ±0.1 °C *** ±0.05 °C ±0.05 °C
  Temperature accuracy (41 ... 90°C)   ±1 °C ±0.1 °C ±0.1 °C ±0.07 °C
  Pressure Range    -700 to +1000 mmHg -600 to +1600 mmHg -700 to +1900 mmHg -700 to +1900 mmHg
  Pressure Accuracy ( 0 ... 300 mmHg)   ±5 mmHg ±1.0 mmHg ± 1 mmHg
± 0.05 mmHg as Option
± 0.5 mmHg
  Pressure Accuracy otherwise   ±1% of reading ±0.5% of reading
+ 1 mmHg
± 2 mmHg ± 1 mmHg
  pH Accuracy   ± 0.1 pH units ± 0.1 pH units ±0.02 pH units ± 0.02 pH units
 
  * In accordance to ISO 10012-1 the accuracy is to low to calibrate any dialysis machine. Read more.
  ** The specified conductivity accuracy by Mesalabs of ±0.10% leads to 0.015 mS/cm at 15 mS/cm.
This seems to be a misprint for following reason.
The reference material with the highes accuracy Mesalab is using comes from
Dansk Fundamental Metrologi A/S.
This material is specified to have a measuring uncertainty of 0.15%. It can be estimated that the calibration procedure at Mesalabs adds another 0.05% uncertainty which leads to an uncertainty of ±0.20%. This uncertainty can roughly estimated to an measuring accuracy of about 0.1%. According to ISO 10012-1 the reference to calibrate a device needs to be at least three times higher than the device to calibrate. Following this rule the calibrated meter has a accuracy of 0.3% which leads to an accuracy of 0.045 mS/cm at 15 mS/cm. The published accuracy by Mesalabs is incorrect and should be about 0.3%.
  *** Due to the temperature compensation the accuracy of 0.1 °C at 37°C leads to a possible error of the conductivity measurement of about 2.4% which computes to 0.36 mS/cm at 15 ms/cm.

 