As option
a higher accuracy is available, which
allows the calibration
of blood
pressure measurement devices in accordance with DIN EN 1060. |
| |
 |
 |
|
For the calibration of blood
pressure measurement devices
we offer a
calibration set according to EN1060.
For more details please refer to "How
to calibrate blood pressure measuring devices?" |
|
Stop Watch |
 |
 |
|
The stop watch function allows the accurate time measurement
with 100
msec. resolution. |
|
Field Calibration |
 |
The
HDM97Pocket
meters can be easily adjusted and calibrated in the field.
The loss of time for calibration and service is minimized. |
|
USB-Interface |
 |
USB 2.0 -
electrically insulated in accordance with ISO 60601-1.
Data exchange with other programs is
possible via the USB-Interface. See also Data
Acquisition |
|
Registration |
|
Manufactured under certified quality
management system in accordance to ISO13485. |
| |
 |
|
Technical Data
|
|
|
| Pressure
|
 |
Reading selectable
as mmHg, kPa, Bar or PSI
 |
 |
|
|
|
Range: |
-700 to + 1900 mmHg,
2,5 Bar |
|
Resolution: |
0.1 mmHg |
|
Accuracy: |
0 ... 300 mmHg ±
1 mmHg
otherwise ±
2 mmHg |
| |
in temperature range 20 to 23 °C
Ceramic sensor fluid resistant |
Option H with higher accuracy |
|
Range: |
-700 to + 1900 mmHg,
2,5 Bar |
|
Resolution: |
0.01 mmHg |
|
Accuracy: |
0 ... 300 mmHg ± 0.5 mmHg,
301 ... 1000 mmHg ± 1 mmHg
otherwise ± 2 mmHg |
| |
in temperature range 20 to 23 °C
Sensor not fluid resistent |
|
|
Power
Supply |
 |
Li-Ion-Polymer battery, rechargeable with
5 V DC, 500 mA
External charger:
Input Voltage 100 … 245 V / 50/60 Hz
Output Voltage 5 V DC, 500 mA
Or via USB-Interface from a computer
Operates more than 200 hours with fully charged battery |
|
Dimensions and Weight |
 |
165 x 82 x 33 mm -
6.5" x 3.2" x 1.3”
0.4 Kg - 0.88 lbs including probe |
|
Sensor connector |
|
 |