Reference Meter for the Service and Calibration of
Medical Devices
Flow
Resistent against dialysate
Innovative Details
Flow Measurement
The sensor allows to calibration of
blood pumps and dialysate flow for example. The reading is shown in ml/min and
the accumulated value.
The sensor allows to connect the dialyser lines of hemodialysis machines.
Stop Watch
The stop watch function allows the accurate time measurement
with 100
msec. resolution.
Field Calibration
The
HDM97Pocket
meters can be easily adjusted and calibrated in the field.
The loss of time for calibration and service is minimized.
USB-Interface
USB 2.0 -
electrically insulated in accordance with ISO 60601-1.
Data exchange with other programs is
possible via the USB-Interface. See also Data
Acquisition
Registration
Manufactured under certified quality
management system in accordance to ISO13485.
Registered as medical device in
USA and Canada.
Technical Data
Flow
Range:
100 ... 2000 ml/min
Resolution:
1 ml/min
Accuracy:
100 ... 500 ml/min ±
1.5 % of upper range value
otherwise ±
2.5 % of upper range value
Power
Supply
Li-Ion-Polymer battery, rechargeable with
5 V DC, 500 mA
External charger:
Input Voltage 100 … 245 V / 50/60 Hz
Output Voltage 5 V DC, 500 mA
Or via USB-Interface from a computer
Operates up to 200 hours with fully charged battery
Dimensions and Weight
165 x 82 x 33 mm -
6.5" x 3.2" x 1.3”
0.4 Kg - 0.88 lbs including probe