What is the
Calibration Period of the IBP Dialysis Meters?
The
calibration period is one year. We recommend to double check the
conductivity once per month with a conductivity reference solution.
If biological
material grows inside the cell it could interfere the accuracy of the
conductivity measurement.
We
know that competing companies do not recommend a monthly test of the
conductivity measurement. On the contrary they offer this as a feature
that their meters needs not to be tested during the calibration
period. From our more than 25 years of experience in developing and
selling dialysis meters we know that this is simply a not true. |